CHAPTER 18 All-on-4 Implants
TECHINIQUE FOR THE ALL-ON-4 CLINICAL SOLUTION
Treatment Planning
General Considerations
1. The surgeon must be sure of the ability to achieve primary implant stability (35 to 45 Ncm insertion torque).
3. The procedure is indicated for a totally edentulous maxilla with a minimum bone width of 5 mm and a minimum bone height of 10 mm from canine to canine.
4. The procedure is indicated for a totally edentulous mandible with a minimum bone width of 5 mm and a minimum bone height of 8 mm between the mental foramina.
7. For tilted posterior implants, the distal screw access holes should be located at the occlusal face of the first molar, the second premolar, or the first premolar (depending on the location of the mental foramen).
NOTE
2. When posterior implants with an internal connection are placed (Fig. 18-1), one of the trichannel lobes on the implant must be pointing distally or slightly buccally.
Specific Considerations: Prosthetics
1. No extensions over one tooth on each side should be made for the immediate all-acrylic bridge, which should have a maximum of 12 teeth.
FLAP SURGICAL APPROACH
Edentulous Mandible
1. Positioning of the All-on-4 Guide (Fig. 18-2)