Fatigue failure of dentin–composite disks subjected to cyclic diametral compression

Highlights

  • Dentin–composite interfaces were evaluated using cyclic diametral compression.

  • Silorane-based systems had higher fatigue resistance than methacrylate-based ones.

  • Reasonable lifetime of restored teeth was predicted by finite element simulation.

Abstract

Objective

Our aim was to establish the relationship between cyclic loading and fatigue life of the dentin–composite interface using the newly developed disk in diametral compression tests. The results were then used to estimate the fatigue life of restored teeth under occlusal loading.

Methods

Disk specimens (5 mm dia. × 2 mm thick) were prepared using bovine incisors and restored with either a methacrylate-based composite Z100™ with Adper Single Bond Plus (Z100) or silorane-based composite Filtek™ LS with LS System adhesive (LS). The dentin–composite disks were tested under cyclic diametral compression to determine the number of cycles to failure ( N f ) at three load levels ( n = 3 per group). Finite element analysis (FEA) was used to calculate the interfacial stresses ( σ ) within the specimen, to establish the σ vs. N f curves, and those within a restored tooth under normal chewing forces (15 N maximum). These were then used to estimate the lifetime of the restored tooth for the two restorative systems.

Results

The disks restored with LS had a higher fatigue resistance than those restored with Z100. The maximum interfacial stress in the restored tooth determined by FEA was ∼0.5 MPa. Based on the estimate of 300,000 cycles of chewing per year, the predicted lifetime under occlusal loading for teeth restored with LS and Z100 was 33 and 10 years, respectively.

Significance

The disk in cyclic diametral compression has been used successfully to provide fatigue data which allows the lifetime of composite-restored teeth under occlusal loading to be predicted using numerical simulation.

Introduction

Resin based dental composites are now widely used for both anterior and posterior restorations. Their comparable mechanical properties to dentin , ability to bond to tooth tissues and superior esthetics make them more favorable than amalgam as a restorative material . Adhesive systems are used to bond the composite to enamel and dentin. For the latter, the impregnation of the demineralized collagen fibrils with a bonding resin is a major factor in the formation of adhesion . While the initial tooth–composite bonds may be adequate, continuous challenges from mastication , hydrolysis and biofilm attacks can degrade the interface, thus resulting in debonding. Secondary caries and fracture of the restoration or tooth, which are the major reasons for the replacement of resin composite restorations, may then follow.

Failures that occur over time as a result of cyclic and/or sustained loading are well recognized clinical problems associated with mineralized tissues . A typical method for quantifying the fatigue behavior of a material is the stress-life approach which relates the number of cycles to failure ( N f ) to the stress amplitude ( σ ). The measured fatigue lifetime represents the number of cycles to cause failure through unstable crack growth. Mineralized tissues such as dentin and enamel display typical σ N f curves in that N f decreases with increasing σ . Clinically, one of the main reasons for the failure of dental restorations is also cyclic fatigue associated with mastication. Being able to quantify the degradation of the tooth–restoration interface in the oral environment can therefore help to predict the lifetime of composite restorations. While the fatigue failure mechanisms of interfaces between composites and tooth tissues are expected to be different from those of the constituents, they display similar σ N f curves. There are relatively few investigations that focus on the cyclic fatigue of the tooth–composite interface , probably because they can be very expensive and time-consuming to perform, even using laboratory testing to simulate the occlusal load. A more efficient strategy, as widely used in engineering, would be to use numerical modeling, coupled with the σ N f data, to predict the lifetimes of dental restorations.

The aims of the present study were, therefore, to (1) determine the σ N f curve, i.e. the fatigue strength, for the composite–dentin interface using the newly developed composite–dentin disk specimen under diametral compression ; and (2) use this interfacial fatigue data to predict the lifetime of composite-restored teeth under physiological conditions. The composite–dentin disk in diametral compression has previously been used successfully to evaluate the static bond strength between these materials .

Materials and methods

Preparation of dentin–composite disks

Dentin–composite disks were prepared using the method reported previously . Briefly, roots from bovine incisors were obtained by removing the crowns at the cemento-enamel junction. They were then trimmed into dentin cylinders of 5 mm in diameter, with the root canals enlarged concentrically to 2 mm in diameter. The dentin cylinders thus produced were randomly divided into two groups and restored with one of two composites (Z100™ and Filtek™ LS, both from 3M ESPE, St. Paul, MN, USA) using the corresponding adhesives as per the manufacturer’s instructions. These two composites were chosen because they have different polymerization chemistries and use different adhesive systems. Z100 is a conventional methacrylate-based composite and it uses a total etch and rinse adhesive system. It exhibits high volumetric shrinkage of over 2% after curing . In contrast, Filtek™ LS is a silorane-based composite with less than 1.0% total volumetric shrinkage , and it uses a self-etch adhesive system. For specimens restored with the Z100 system, the inner dentin surface was etched with 35% phosphoric acid for 20 s and then rinsed with deionized water. Two layers of adhesive (Adper™ Single Bond Plus, 3M ESPE) were applied to the etched surface and cured for 20 s. To minimize shrinkage stress, Z100 composite was applied incrementally to fill the cylinders. Each increment, less than 2-mm thick, was cured for 40 s to ensure adequate curing. For specimens restored with the LS system, a layer of Self-Etch Primer (LS System Adhesive, 3M ESPE) was first applied to the inner dentin surface and cured for 10 s. This was followed by the application of a layer of Bond (LS System Adhesive, 3M ESPE) with 20 s of curing. LS composite was then applied incrementally and cured in the same way as Z100. Finally, the composite-restored cylinders were transversely cut to produce 2-mm thick round disks. All restorative materials were cured using a LED light-curing unit (Elipar™ S10, 3M ESPE) with a power density of 1200 mW/cm 2 . During curing, the tip of light guide was held as close as possible to the materials without touching them. The compositions and product information of the composites and adhesive bonding systems are described in Table 1 .

Table 1
Product information of composites and adhesives used in the study.
Materials Compositions Lot # Expiration date
Z100™ Bis-GMA, TEGDMA, silane treated microfiller (66 vol%), initiator N362979 2014-12
Filtek™ LS Silorane resin, silane treated quartz filler (53 vol%), yttrium floride, initiator, stabilizers N554531 2015-05
Adper™ Single Bond Plus Etch-and-rinse, conventional adhesive system; Bis-GMA; HEMA, Glycerol 1,3-dimethacrylate, copolymer of acrylic and itaconic acids, UDMA, initiator, ethanol, water N437332 2015-10
LS System adhesive Two-bottle self-etch adhesive system; Primer: Bis-GMA, HEMA, phosphorylated methacrylates, 1,6-hexanediol dimethacrylate, copolymer of acrylic and itaconic acid, silane-treated silica, initiators, stabilizers, water, ethanol. Bond: substituted dimethacrylate, TEGDMA, phosphorylated methacrylate, 1,6-hexanediol dimethacrylate, silane-treated silica filler, initiators, stabilizers N522061 2015-05

Fatigue testing

The prepared disks were loaded to failure in cyclic diametral compression using a universal testing system (MTS 858 Mini Bionix II, Eden Prairie, MN, USA) operated with load control ( Fig. 1 a) . Cyclic diametral compression at 1 Hz with a square wave of zero minimum load was applied to the specimens in deionized water at room temperature. The load and displacement were monitored continuously throughout the whole cyclic loading process. The number of cycles to failure ( N f ) was determined from a dramatic change in the displacement patterns ( Fig. 1 b). The static fracture load of both Z100- and LS-restored disks had previously been found to be approximately 400 N . For the current fatigue test, three load levels, with 50%, 37.5% and 25% of the static fracture load as the peak value, were applied, and three samples for each material group were evaluated at each load level. The use of such a small sample size was justified because the uncertainties in the parameters derived for the fatigue models were reasonably small (see standard errors in Fig. 6 ). The stress-life ( σ N f ) curves were constructed by first determining the maximum interfacial tensile stress within the disk using finite element analysis (FEA, see later) with the peak load and then plotting the stress against the number of cycles to failure.

Fig. 1
(a) Experimental set-up for the diametral compression fatigue test. (b) Typical patterns of axial stroke and axial load as a function of time. (c) Patterns of axial stroke and load from 650 s to 660 s of the test in (b) showing constant-amplitude cyclic loading and displacement. (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

Linear regression was performed on the σ N f data using appropriate mathematical models for each restorative system. The quality of the linear regression and the uncertainties in the parameters for the fatigue models were then assessed using statistical analysis.

Scanning electron microscopy (SEM)

SEM was used to examine the morphology and fracture modes of the dentin–composite disks after fatigue testing. The fractured disks were air-dried overnight and mounted on aluminum stubs with carbon tapes. They were examined by a semi-environmental tabletop SEM (TM-3000, Hitachi, Japan) operated at a 15-kV accelerating voltage. No conductive coating was applied to the samples. To counteract the charging effect, the charge-up reduction mode was used for acquiring images. The working distance was in the range of 5–8 mm. The failure modes were classified as (1) cohesive failure in the resin composite; (2) interfacial failure between the adhesive and composite; (3) cohesive failure in the adhesive; or (4) interfacial failure between the adhesive and dentin.

Finite element analysis of interfacial stresses

Stresses at the dentin–composite interface of the disk under diametral compression were calculated with FEA for different load levels. The model of the specimen, together with the two horizontal steel plates for load application ( Fig. 2 a) , was constructed using Hypermesh 11.0 (HyperWorks, Altair Engineering, USA). The frictional coefficient between the loading plates and the specimen was set as 0.3 . Each component was assumed to be isotropic, homogeneous and linear elastic. Table 2 lists the elastic modulus and Poisson’s ratio of dentin, enamel, steel, Z100 and LS. The disk model was analyzed using the FEA software ABAQUS 6.11 (SIMULIA, Providence, RI, USA). The maximum tensile interfacial stress was determined as a function of the applied load, which allowed the experimental data (load vs. N f ) to be presented as σ N f curves. In addition, the maximum tensile stress in the dentin was also calculated and compared with its reported fracture strength to assess the possibility that dentin cohesive fracture preceded interfacial fracture.

Fig. 2
(a) Principal stress distribution in the dentin–composite disk. (b) Contact pressures at the disk composite–dentin bonding interface with blue-colored areas under tensile stresses (negative pressure) and red-colored areas under compressive stress. (c) Maximum interfacial tensile stresses in disks restored with LS and Z100 as a function of load. (d) The maximum tensile stress in the dentin near the loading point as a function of load. (e) Finite element model of two opposing molars. The mandibular molar was subjected to occlusal (vertical) force of 15 N and lateral (horizontal) displacement of 0.2 mm to simulate chewing movement. (f) Distribution of the maximum principal stress in the restored tooth. (g) and (h) Contact pressure at the tooth–restoration interface with the same color convention as in (b). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

Table 2
Materials properties for finite elemental analysis.
Material Elastic modulus (GPa) Poisson’s ratio
Dentin 18.6 0.31
Enamel 84 0.3
Z100 8.5 0.3
LS 13.5 0.3
Steel 210 0.3

As an example to demonstrate the application of an engineering approach to assessing the fatigue lifetime of composite restorations, the stresses of a tooth with a mesial–occlusal–distal (MOD) restoration under normal chewing motions were calculated using FEA ( Fig. 2 e). Failure of the restored tooth was assumed to be caused by debonding at the dentin–composite interface. A 2nd mandibular molar and a 2nd maxillary molar were scanned with an X-ray micro-computed tomography (micro-CT) machine (XT H 225, Nikon Metrology, United Kingdom) using the following operational parameters: 90 kV, 90 μA, 720 projections and 4 frames per projection. Reconstructions of the objects in 3D space were done with CT Pro 3D (Nikon Metrology, Brighton, MI, USA). The 3D CT images were used to build finite element models for the teeth. First, the surfaces of the dentin and enamel of each tooth were extracted using Amira (Mercury Computer Systems, Berlin, Germany). Solid tetrahedral elements were then created inside the volumes bounded by the surfaces using Hypermesh. A MOD restoration was introduced into the mandibular molar model, as shown in Fig. 2 e. These two tooth models were first placed in the cusp-against-cusp position, simulating initial occlusion. The root of the maxillary molar was fully fixed, while the root of the mandibular molar was constrained in the mesial–distal direction and rotationally about all 3 axes, thus restricting its motions along the vertical and buccal–lingual directions only. A normal chewing or mastication cycle is composed of opening, closing and occlusion where the occlusal phase ends ∼0.5 mm from the maximum intercuspation position . The chewing motion was thus simulated with 2 load steps. In the first load step, an upward occlusal force of 15 N was applied to the mandibular molar. In the second load step, a lateral displacement of 0.2 mm was applied to the mandibular molar to bring it to centric occlusion. The model was again analyzed using ABAQUS to obtain the maximum interfacial tensile stress, which was then used, in conjunction with the experimental σ N f curves, to estimate the fatigue lifetime of the restored tooth under occlusal loads.

Materials and methods

Preparation of dentin–composite disks

Dentin–composite disks were prepared using the method reported previously . Briefly, roots from bovine incisors were obtained by removing the crowns at the cemento-enamel junction. They were then trimmed into dentin cylinders of 5 mm in diameter, with the root canals enlarged concentrically to 2 mm in diameter. The dentin cylinders thus produced were randomly divided into two groups and restored with one of two composites (Z100™ and Filtek™ LS, both from 3M ESPE, St. Paul, MN, USA) using the corresponding adhesives as per the manufacturer’s instructions. These two composites were chosen because they have different polymerization chemistries and use different adhesive systems. Z100 is a conventional methacrylate-based composite and it uses a total etch and rinse adhesive system. It exhibits high volumetric shrinkage of over 2% after curing . In contrast, Filtek™ LS is a silorane-based composite with less than 1.0% total volumetric shrinkage , and it uses a self-etch adhesive system. For specimens restored with the Z100 system, the inner dentin surface was etched with 35% phosphoric acid for 20 s and then rinsed with deionized water. Two layers of adhesive (Adper™ Single Bond Plus, 3M ESPE) were applied to the etched surface and cured for 20 s. To minimize shrinkage stress, Z100 composite was applied incrementally to fill the cylinders. Each increment, less than 2-mm thick, was cured for 40 s to ensure adequate curing. For specimens restored with the LS system, a layer of Self-Etch Primer (LS System Adhesive, 3M ESPE) was first applied to the inner dentin surface and cured for 10 s. This was followed by the application of a layer of Bond (LS System Adhesive, 3M ESPE) with 20 s of curing. LS composite was then applied incrementally and cured in the same way as Z100. Finally, the composite-restored cylinders were transversely cut to produce 2-mm thick round disks. All restorative materials were cured using a LED light-curing unit (Elipar™ S10, 3M ESPE) with a power density of 1200 mW/cm 2 . During curing, the tip of light guide was held as close as possible to the materials without touching them. The compositions and product information of the composites and adhesive bonding systems are described in Table 1 .

Table 1
Product information of composites and adhesives used in the study.
Materials Compositions Lot # Expiration date
Z100™ Bis-GMA, TEGDMA, silane treated microfiller (66 vol%), initiator N362979 2014-12
Filtek™ LS Silorane resin, silane treated quartz filler (53 vol%), yttrium floride, initiator, stabilizers N554531 2015-05
Adper™ Single Bond Plus Etch-and-rinse, conventional adhesive system; Bis-GMA; HEMA, Glycerol 1,3-dimethacrylate, copolymer of acrylic and itaconic acids, UDMA, initiator, ethanol, water N437332 2015-10
LS System adhesive Two-bottle self-etch adhesive system; Primer: Bis-GMA, HEMA, phosphorylated methacrylates, 1,6-hexanediol dimethacrylate, copolymer of acrylic and itaconic acid, silane-treated silica, initiators, stabilizers, water, ethanol. Bond: substituted dimethacrylate, TEGDMA, phosphorylated methacrylate, 1,6-hexanediol dimethacrylate, silane-treated silica filler, initiators, stabilizers N522061 2015-05

Fatigue testing

The prepared disks were loaded to failure in cyclic diametral compression using a universal testing system (MTS 858 Mini Bionix II, Eden Prairie, MN, USA) operated with load control ( Fig. 1 a) . Cyclic diametral compression at 1 Hz with a square wave of zero minimum load was applied to the specimens in deionized water at room temperature. The load and displacement were monitored continuously throughout the whole cyclic loading process. The number of cycles to failure ( N f ) was determined from a dramatic change in the displacement patterns ( Fig. 1 b). The static fracture load of both Z100- and LS-restored disks had previously been found to be approximately 400 N . For the current fatigue test, three load levels, with 50%, 37.5% and 25% of the static fracture load as the peak value, were applied, and three samples for each material group were evaluated at each load level. The use of such a small sample size was justified because the uncertainties in the parameters derived for the fatigue models were reasonably small (see standard errors in Fig. 6 ). The stress-life ( σ N f ) curves were constructed by first determining the maximum interfacial tensile stress within the disk using finite element analysis (FEA, see later) with the peak load and then plotting the stress against the number of cycles to failure.

Fig. 1
(a) Experimental set-up for the diametral compression fatigue test. (b) Typical patterns of axial stroke and axial load as a function of time. (c) Patterns of axial stroke and load from 650 s to 660 s of the test in (b) showing constant-amplitude cyclic loading and displacement. (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

Linear regression was performed on the σ N f data using appropriate mathematical models for each restorative system. The quality of the linear regression and the uncertainties in the parameters for the fatigue models were then assessed using statistical analysis.

Scanning electron microscopy (SEM)

SEM was used to examine the morphology and fracture modes of the dentin–composite disks after fatigue testing. The fractured disks were air-dried overnight and mounted on aluminum stubs with carbon tapes. They were examined by a semi-environmental tabletop SEM (TM-3000, Hitachi, Japan) operated at a 15-kV accelerating voltage. No conductive coating was applied to the samples. To counteract the charging effect, the charge-up reduction mode was used for acquiring images. The working distance was in the range of 5–8 mm. The failure modes were classified as (1) cohesive failure in the resin composite; (2) interfacial failure between the adhesive and composite; (3) cohesive failure in the adhesive; or (4) interfacial failure between the adhesive and dentin.

Finite element analysis of interfacial stresses

Stresses at the dentin–composite interface of the disk under diametral compression were calculated with FEA for different load levels. The model of the specimen, together with the two horizontal steel plates for load application ( Fig. 2 a) , was constructed using Hypermesh 11.0 (HyperWorks, Altair Engineering, USA). The frictional coefficient between the loading plates and the specimen was set as 0.3 . Each component was assumed to be isotropic, homogeneous and linear elastic. Table 2 lists the elastic modulus and Poisson’s ratio of dentin, enamel, steel, Z100 and LS. The disk model was analyzed using the FEA software ABAQUS 6.11 (SIMULIA, Providence, RI, USA). The maximum tensile interfacial stress was determined as a function of the applied load, which allowed the experimental data (load vs. N f ) to be presented as σ N f curves. In addition, the maximum tensile stress in the dentin was also calculated and compared with its reported fracture strength to assess the possibility that dentin cohesive fracture preceded interfacial fracture.

Nov 23, 2017 | Posted by in Dental Materials | Comments Off on Fatigue failure of dentin–composite disks subjected to cyclic diametral compression