Chapter 14 Assessment of the Dentition
Procedure 14-1 USE OF AN ELECTRIC PULP TESTER TO DETERMINE PULP VITALITY
EQUIPMENT | |
Personal protective equipment | |
2 × 2 gauze | |
Saliva ejector | |
Cotton rolls | |
Toothpaste | |
Electric pulp tester | |
STEPS | RATIONALES |
1. Assemble equipment. | Manufacturers’ instructions must be followed. When the tester rheostat is separate from the applicator tip, an assistant is needed. |
2. Review health history. | Use of the electric pulp tester is contraindicated for a client with a pacemaker or any electronic life-support device because it can interfere with the device’s function and pose a serious health threat. |
3. Describe the test’s purpose and methods. | This familiarizes the client with the procedure. For some pulp testers, the client lightly holds the handle with the gloved clinician to complete the circuit. |
4. Explain that the client may feel a tingling or a warm sensation. | |
5. Identify the suspected tooth and a “cont/> |