Adhesion of living cells to abutment materials, dentin, and adhesive luting cement with different surface qualities

Highlights

  • A new testing method for researching real “cell-attachment” of living cells is proposed.

  • Human dentin could be established as a gold standard for this testing method.

  • Zirconium dioxide, titanium and lithium disilicate are comparable abutment materials with regard to cell adhesion.

  • Machined – not polished or roughened – abutment surfaces provide the best cell attachment.

  • The resin cement of two-piece abutments does not pose a problem for cell adhesion.

Abstract

Objective

We tested the adhesion properties of living gingival fibroblasts on three different implant abutment materials, adhesive resin used to bond bi-partite abutments, and human dentin.

Methods

Discs of lithium disilicate (LS), zirconium dioxide (Zr), adhesive resin cement (AR), titanium (Ti), and human dentin (HD) were fabricated with three different levels of surface roughness (rough, machined, and polished). Ra and Rz, water contact angle, and cell detachment forces were measured. Cell detachment force was measured for single cells using single-cell force spectroscopy. Data were statistically analyzed using parametric tests (ANOVA, MANOVA, Bonferroni post-hoc tests).

Results

Surface roughness significantly influenced the water contact angle for all materials (P ≤ 0.05). Overall, HD showed the lowest contact angle, followed by LS, Ti, Zr, and AR (P ≤ 0.05). Comparison of cell detachment forces between materials with rough and machined surfaces revealed no significant differences (P > 0.05), with the exception of Zr compared to HD with rough surfaces (P = 0.006). For polished surfaces, HD showed the highest detachment force (P ≤ 0.0001), followed by Ti, AR, and Zr, which did not significantly differ from each other (P > 0.05) and LS; Ti/AR was significantly different from LS (P ≤ 0.05). Except for HD, where polished surfaces exhibited the highest cell detachment force (P ≤ 0.002), most machined surfaces showed higher cell detachment forces than polished or rough surfaces.

Significance

Implant abutments should ideally be provided with a machined like surface roughness for best cell adhesion.

Introduction

Dental implant abutments should have the following characteristics: (1) ensure stability of both, soft and hard tissues ; (2) cause no, or only limited, wear of the implant/abutment connection ; (3) exhibit low plaque accumulation rates ; (4) be biocompatible (no allergic reactions, no stimulation of the immune system ; and (5) should be aesthetically acceptable (no metal discolouration of the gingiva) . Commonly used materials providing most, but not all, of these properties include: CP4 pure titanium Grade 1 abutments , zirconium dioxide ceramic abutments , and recently introduced lithium disilicate ceramic abutments .

Marginal bone loss of approximately 0.6 mm appears to be inevitable because of surgical trauma after implant placement . Further bone loss should be avoided to prevent aesthetical problems . The established peri-implant connective tissue resembles scar tissue in composition, fibre orientation, and vasculature . The connective tissue forms a ring around the abutment and rests on the platform-switched implant shoulders. One of the main functions of the peri-implant soft tissue is to seal the marginal bone and protect the host from bacterial penetration .

Surface properties modify cell behaviour , and surface alterations can achieve specific cell responses . Although many parameters are involved, surface roughness and wettability are two important factors for cell adhesion and proliferation .

Previous studies have improved the implant seal using different cleaning procedures or surface alterations . The abutment materials evaluated in most of these studies were titanium abutments , with one study using zirconium dioxide and another using lithium disilicate . It has generally been established that a rough titanium surface increases cell growth/migration and expression of proteins in the extracellular matrix . Although both, osseous- and soft tissue healing around dental implants are critical for clinical success, there is limited information available regarding the effect of different implant surfaces on cell adhesion and integrin expression in soft tissues . Most studies have relied on percentage-wise anatomic descriptions of cell proportions and cell growth analysis . Some studies have investigated the biochemical and physiological details of cell alterations and adhesion properties . In recent years, single-cell force spectroscopy has become a valuable tool for investigating the adhesion properties of cells , particularly because it enables comparison of adhesion on different surfaces using the same cell .

The null-hypotheses to be tested were:

Short-term cell adhesion and water contact angle measurements using single-cell force spectroscopy are not influenced by the use of three different implant abutment materials, adhesive resin and dentin. Equally, different surface roughnesses have no significant influence on short-term cell adhesion and water contact angle measurements.

Materials and methods

Materials used

The main materials used in this study are shown in Table 1 . Overall, five materials with three different surface characteristics were tested: lithium disilicate (LS), zirconium dioxide (Zr), adhesive resin cement (AR), titanium (Ti), and human dentin (HD). Specimens of LS (IPS e.max CAD MO3, Camlog Dedicam, Wimsheim, Germany) composed of SiO 2 (57–80%; all shown in weight%), Li 2 O (11–19%), KO 2 (0–13%), P 2 O 5 (0–11%), ZrO 2 (0–8%), ZnO (0–8%), and others oxides and pigments (around 7%). Samples of Zr (Zirlux FC2 U5, Camlog Dedicam) composed of ZrO 2 (>91%), Y 2 O 3 (5.3%), Al 2 O 3 (<0.15), and HfO 2 (<3%). Specimens of AR (Multilink Hybrid Abutment, Ivoclar Vivadent, Schaan, Liechtenstein) composed of (Base) Yb 3 F 3 (10–25%), ethoxylated bisphenol-a-dimethacrylate (10–25%), Bis-GMA (10–25%), 2-hydroxyethyl-methacrylate (3–10%), and 2-dimethylaminoethylmethacrylate (0.3–1%) and (Catalyst) Yb 3 F 3 (10–25%), ethoxylated bisphenol-a-dimethacrylate (10–25%), urethane dimethacrylate (3–10%), 2-hydroxyethyl-methacrylate (3–10%), and dibenzoylperoxid (1–2.5%). Specimens of Ti (Ti-6Al-4V ELI (ASTM F 136), Camlog Dedicam) consisted of Al (5.5–6.5%), V (3.5–4.5%), and C, Fe, O, N, and H (all <0.13%) with the remaining amount comprised of titanium.

Table 1
Main materials used in this study. Data listed below are provided by the manufacturers.
Proprietary material Lot no. Type Manufacturer
Multilink Hybrid Abutment T15412 Adhesive resin Ivoclar Vivadent, Schaan, Liechtenstein
Titanium Ti-6Al-4V ELI D0065.6305 Titanium alloy Camlog Dedicam, Wimsheim, Germany
Lithium disilicate D0065.6408 IPS e.max CAD MO3 Camlog Dedicam
Zirconium dioxide D0065.6347 Zirlux FC2 U5 Camlog Dedicam
C2242.4308
Human gingiva fibroblast cells 305110-212 Fibroblast cells Cell line services Eppelheim, Germany

Specimen preparation

LS, Zr, and Ti specimens were manufactured by CAD/CAM (Camlog Dedicam). The dimensions of the specimens were 10 mm (depth) × 10 mm (width) × 1 mm (height), with the exception of HD, which naturally exhibits minor irregularities in depth and width. The ceramics were used in the sintered stage. Overall, 24 specimens per group were prepared.

AR specimens were produced by initially taking impressions of Ti specimens (Gammasil Perfect TEC A85, Müller-Omicron, Cologne, Germany). Next, AR was mixed according to the manufacturer’s instructions, placed into the silicone mould, and covered with glycerine gel (Liquid Strip, Ivoclar Vivadent, Schaan, Liechtenstein). After 10 min, the AR block was placed in de-ionized water in a heated chamber at 37 °C (99 °F) for 24 h.

In order to prepare the HD specimens, 24 intact non-carious human third molars were extracted and stored in 0.1% thymol-saturated isotonic saline solution at 4 °C. The teeth were equally divided for all experimental groups. A 2-mm-thick dentin disc was sliced perpendicularly to the longitudinal axis of each tooth using a slow-speed diamond saw (Isomet 1000, Buehler, Lake Bluff, IL, USA) under water cooling. The resulting flat dentin surface was evaluated using a stereomicroscope (Wild Makroskop M420, Heerbrugg, Switzerland) at 10-fold magnification to confirm the absence of enamel at the occlusal dentin surface. The dentin was maintained under wet conditions during all preparations by storing it in de-ionized water.

Preparation of surface roughness

Three different surface qualities, rough, machined, and polished, were tested in this study. Eight specimens per material and group were prepared (for detailed information, see Table 2 ).

Table 2
Preparation of different surface finishes. Polished surface treatment was followed by the preparation of machined surfaces.
Material Grinding medium Grinding pressure in N Grinding time in seconds
Rough surface finish
LS P 600 Silicon carbide abrasive paper, air abrasion with Al 2 O 3 20 30
Zr P 600 Diamond disc (Apex DGD, 35 ηm, Φ 254 mm (10″), Buehler, Lake Bluff, IL, USA), air abrasion with Al 2 O 3 20 120
AR P 600 Silicon carbide abrasive paper, air abrasion with Al 2 O 3 Pressure manually applied 10
Ti P 600 Silicon carbide abrasive paper, air abrasion with Al 2 O 3 20 60
HD P 600 Silicon carbide abrasive paper, air abrasion with Al 2 O 3 20 15
Machined surface finish
LS P 1200 Silicon carbide abrasive paper 20 N 60
Zr Manually with a handpiece (blue polisher, Komet Dental, Lemgo, Germany), no water cooling Pressure manually applied at 600 revs/m 30
AR 1. P 1200 Silicon carbide abrasive paper 1. Pressure manually applied 1. 20
2. P 2500 Silicon carbide abrasive paper 2. Pressure manually applied 2. 30
3. Yellow Silicon polisher (Henry Schein Dental, Melville, NY, USA) 3. Pressure manually applied at 700 revs/m 3. 30
Ti P 1200 Silicon carbide abrasive paper 20 N 120
HD P 1200 Silicon carbide abrasive paper 20 N 30
Polished surface finish
LS P 2500 Silicon carbide abrasive paper 20 N 120
Zr Manually with a handpiece (grey polisher, Komet Dental, Lemgo, Germany), no water cooling Pressure manually applied at 600 revs/m 30
AR Wool high polisher hard and soft (Henry Schein Dental, Melville, NY, USA), no water cooling Pressure manually applied at 1000 revs/m 30
Ti P 2400 Silicon carbide abrasive paper 20 N 240
HD 1. P 2400 Silicon carbide abrasive paper 1. 20 N 1. 40
2. P 4000 Silicon carbide abrasive paper 2. 20 N 2. 90
Only gold members can continue reading. Log In or Register to continue

Stay updated, free dental videos. Join our Telegram channel

Nov 23, 2017 | Posted by in Dental Materials | Comments Off on Adhesion of living cells to abutment materials, dentin, and adhesive luting cement with different surface qualities

VIDEdental - Online dental courses

Get VIDEdental app for watching clinical videos